Cavity ring-down spectroscopy for detection in liquid chromatography at UV wavelengths using standard cuvettes in a normal incidence geometry

L. van der Sneppen, F. Ariese, C. Gooijer, W.M.G. Ubachs

Research output: Contribution to JournalArticleAcademicpeer-review

Abstract

Liquid chromatography (LC) with cavity ring-down spectroscopy (CRDS) detection, using flow cuvettes (put under normal incidence inside the ring-down cavity), is demonstrated. Fresnel reflections are maintained within the capture range of a stable cavity of 4 cm length. This method circumvents the need for specific Brewster's angles and possible mirror degradation is avoided. The flow cuvettes are commercially available at low cost. At 355 nm (the frequency-tripled output of a Nd:YAG laser), the system surpasses the performance of conventional absorbance detectors; the baseline noise was 1.3 × 10
Original languageEnglish
Pages (from-to)184-8
JournalJournal of Chromatography A
Volume1148
Issue number2
DOIs
Publication statusPublished - 2007